Chapter 4 Calibration and Adjustment
Performance Verification Tests
4
Performance Verification Tests
Use the Performance Verification Tests to verify the measurement
performance of the instrument. The performance verification tests use
the instrument’s specifications listed in the “Specifications” chapter
beginning on page 17.
You can perform three different levels of performance verification tests:
Self-Test A series of internal verification tests that give high
confidence that the instrument is operational.
Quick Verification A combination of the internal self-tests and
selected verification tests.
Performance Verification Tests An extensive set of tests that are
recommended as an acceptance test when you first receive the
instrument or after performing adjustments.