93
Chapter 4 Calibration and Adjustment
Performance Verification Tests
4
4
Self-Test
A brief power-on self-test occurs automatically whenever you turn on the
instrument. This limited test assures that the instrument is operational.
To perform a complete self-test:
1 Press System Key on the front panel.
2 Select the Self Test softkey from the “Utility > Test/Cal” menu.
A complete description of the self-tests can be found in chapter 6.
The instrument will automatically perform the complete self-test
procedure when you release the key. The self-test will complete in
approximately 15 seconds.
If the self-test is successful, “Self Test Passed” is displayed on the
front panel.
If the self-test fails, “Self Test Failed” and the number of errors are
displayed. The errors are viewable under the “Help” softkey.
If repair is required, see chapter 6, “Service,” for further details.