EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>33522A

Agilent Technologies 33522A Service Guide

Agilent Technologies 33522A
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #111 background imageLoading...
Page #111 background image
111
Chapter 4 Calibration and Adjustment
Internal ADC Adjustment
4
4
Internal ADC Adjustment
The waveform generator stores calibration constants related to the gain
and offset of the internal ADC. Setup 6 must always be performed
before any other amplitude adjustments are attempted. The internal
ADC is then used as a source for the calibration constants generated in
self calibration (setup 7).
1 Connect the waveform generator Channel 1 Output to the waveform
generator rear panel Modulation Input and DMM as shown below.
2 Set the DMM to display 5 1/2 digits and function to dcV.
3 Enter the following setup.
* Constants are stored after completing this setup.
4 Use the numeric keypad or knob to enter the value measured on the
DMM.
Note This setup requires approximately 15 seconds to complete.
5 Disconnect all cables from the instrument
Nominal Signal
Setup DC level
6* ~1.0 Vdc ±10% Calibrates the internal ADC.
DMM
Modulation In
Front Panel
Rear Panel

Table of Contents

Other manuals for Agilent Technologies 33522A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33522A and is the answer not in the manual?

Agilent Technologies 33522A Specifications

General IconGeneral
BrandAgilent Technologies
Model33522A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals