Chapter 4 Calibration and Adjustment
Self Calibration Adjustment (Channel 2)
4
Self Calibration Adjustment (Channel 2)
1 Enter and begin the following setup.
* Constants are stored after completing this setup.
2 After performing setup 54:
a. If your calibration procedures require you to verify the adjustment
just made, exit the calibration menu and perform “DC Offset
Voltage Verification”, on page 98.
b. If you are making all the adjustments and then verifying the
instrument’s performance, continue with the next procedure in
this chapter.
Note This setup requires approximately 15 seconds to complete.
Setup
54* Self-calibration. The output is disabled.