EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>33522A

Agilent Technologies 33522A Service Guide

Agilent Technologies 33522A
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #113 background imageLoading...
Page #113 background image
113
Chapter 4 Calibration and Adjustment
Output Impedance Adjustment
4
4
Output Impedance Adjustment
The waveform generator stores calibration constants for the output
Impedance of the instrument. The output impedance constants are
generated with and without the post-amplifier attenuator.
1 Set the DMM to measure offset-compensated, four-wire Ohms. Set the
DMM to use 100 NPLC integration. Connect the Ohms Source and Ohms
Sense DMM inputs to the Channel 1 output as shown below.
2 Use the DMM to make a 4-wire resistance measurement at the front
panel Output connector for each setup in the following table. The
expected measured value is approximately 50 .
* Constants are stored after completing this setup.
3 Using the numeric keypad or knob, adjust the displayed impedance at
each setup to match the measured impedance. Select ENTER VALUE.
4 There are no specific operational verification tests for Output
Impedance. Continue with the next adjustment procedure in this
chapter.
Setup
8* -24 dB post-attenuator range
9* 0 dB range
DMM

Table of Contents

Other manuals for Agilent Technologies 33522A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33522A and is the answer not in the manual?

Agilent Technologies 33522A Specifications

General IconGeneral
BrandAgilent Technologies
Model33522A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals