Chapter 4 Calibration and Adjustment
-24 dB Range Flatness Adjustment (Channel 2)
4
4
3 Using the numeric keypad or knob, adjust the displayed voltage at each setup
to match the measured voltage. Select ENTER VALUE.
4 After performing setup 92:
a. If your calibration procedures require you to verify the adjustment
just made, exit the calibration menu and perform “-24 dB Range
Flatness Verification”, on page 101 for Channel 2.
b. If you are making all the adjustments and then verifying the
instrument’s performance, continue with the next procedure in
this chapter.