Chapter 4 Calibration and Adjustment
-8 dB Range Flatness Adjustment (Channel 2)
4
4
3 Using the numeric keypad or knob, adjust the displayed voltage at each setup
to match the measured voltage. Select ENTER VALUE.
4 After performing setup 100, you have now completed the recommended
adjustment procedures. Verification of the output specifications is
recommended.
a. If your calibration procedures require you to verify the adjustment
just made, exit the calibration menu and perform “-8 dB Range
Flatness Verification”, on page 99.