EasyManuals Logo

Agilent Technologies 34970A Service Guide

Agilent Technologies 34970A
273 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #169 background imageLoading...
Page #169 background image
Self-Test Procedures
Power-On Self-Test
Each time the instrument is powered on, a small set of self-tests are
performed. These tests check that the minimum set of logic and
measurement hardware are functioning properly. Any plug-in
modules installed are verified for two-way communication with the
main controller.
Complete Self-Test
Hold down any front panel key for 5 seconds while turning on the
power to perform a complete self-test. The instrument beeps when
the test starts. If all self-tests pass the display shows PASS for
five seconds and the instrument returns to the last measurement
function.
Plug-in Module Self-Test
No user self-test exists for the plug-in modules. The plug-in modules
perform their own self-test when power is applied. Additionally, the
mainframe checks two-way communication with all plug-in modules
at power on.
Each plug-in module also performs error checking at regular
intervals during operation and any errors detected are reported via
the status system to the main controller.
Note: The following pages contain a subset of the 34970A error
messages. Refer to Chapter 6 in the 34970A User’s Guide (p/n
34970-90003) for the complete error message listing.
6
Chapter 6 Service
Self-Test Procedures
167

Table of Contents

Other manuals for Agilent Technologies 34970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34970A and is the answer not in the manual?

Agilent Technologies 34970A Specifications

General IconGeneral
BrandAgilent Technologies
Model34970A
CategorySwitch
LanguageEnglish

Related product manuals