EasyManuals Logo

Agilent Technologies 34970A Service Guide

Agilent Technologies 34970A
273 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #87 background imageLoading...
Page #87 background image
–10 Vdc Adjustment Procedure (Optional)
The –10 Vdc calibration electronically enhances the Internal DMM’s
a-to-d converter linearity characteristic. This adjustment should
ONLY be performed after servicing the A-to-D converter or
replacement of the calibration RAM.
You will need a 34901A 20-Channel Multiplexer to perform the
following procedures (see page 66). Install the Multiplexer in slot 200.
1 If a zero calibration has not been performed recently, perform one
before beginning this procedure (see page 81).
2 Select Channel 210. Configure the channel as follows:
DC VOLTS
10 V range
6
1
2
digits
INTEG 100 PLC (in the Advanced menu)
INPUT R > 10 G (in the Advanced menu)
Before executing each test, you must press to enable reading
monitoring on the selected channel (or use the ROUTe:MON command
from the remote interface).
3 Measure and note the voltage offset present at the end of the
measurement cable by shorting the ends of the Channel 210
measurement cable. Be sure to use a copper wire and allow enough
time for the residual thermal offset to stabilize (usually about
1 minute).
4 Connect the input cable to the calibrator output and set the calibrator
to output +10V. Allow enough settling time for any thermal offset
voltages to stabilize (usually about 1 minute).
5
Perform a +10V dc gain calibration. Press to enter the
calibration menu. Press again to begin the adjustment procedure.
6
Use the knob, and to set the number in the display to the
sum of the calibrator output and the measured offset (from step 3)
and press . For example, if the calibrator output is 10.001 volts
and the measured offset is 10 µV, enter +10.001010 volts. When the
adjustment finishes, verify that new readings fall within ± 20 µV of
the calibrator output plus the offset.
Continued on next page...
4
Chapter 4 Calibration Procedures
–10 Vdc Adjustment Procedure (Optional)
85

Table of Contents

Other manuals for Agilent Technologies 34970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34970A and is the answer not in the manual?

Agilent Technologies 34970A Specifications

General IconGeneral
BrandAgilent Technologies
Model34970A
CategorySwitch
LanguageEnglish

Related product manuals