EasyManua.ls Logo

Agilent Technologies 34970A - Calibration Procedure; Aborting a Calibration in Progress

Agilent Technologies 34970A
273 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Procedure
The following procedure is the recommended method to complete an
instrument calibration.
1 Read “Test Considerations” (page 71).
2 Unsecure the instrument for calibration (page 51).
3 Perform the verification tests to characterize the instrument
(incoming data).
4 Perform the zero adjustment procedures.
5 Perform the gain adjustment procedures. Perform the verification
tests to verify the adjustments (outgoing data).
6 Secure the instrument against calibration.
7 Note the new security code and calibration count in the
instrument’s maintenance records.
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the
procedure has already been initiated. You can abort a calibration at
any time by turning off the power. When performing a calibration
from the remote interface, you can abort a calibration by issuing a
remote interface device clear message.
CAUTION
If you abort a calibration in progress when the instrument is attempting
to write new calibration constants to EEPROM, you may lose all
calibration constants for the function. Typically, upon re-applying
power, the instrument will report error 705 Cal:Aborted. You may
also generate errors 740 through 746. If this occurs, you should not use
the instrument until a complete re-adjustment has been performed.
Chapter 4 Calibration Procedures
Calibration Procedure
70

Table of Contents

Other manuals for Agilent Technologies 34970A

Related product manuals