Chapter 2 Operating Your Frequency Counter
Using the LIMITS Menu Keys
2-34 Operating Guide
2
Limits Testing Example 2—Flag On Limits but Continue Measuring
If you decide that you want the Counter to flag measurements that are out of limits but
to continue taking measurements, then perform the following steps.
1 Press Limit Modes key until ON FAIL: STOP is displayed.
2 Press any one of the arrow keys to toggle to the next state.
ON FAIL: GO ON is displayed.
3 Press Run key to start making measurements with this limits-testing choice
selected.
Now, each time a measurement drifts out of the user-entered limits the Limit
annunciator in the display will light, but the Counter will continue to make
measurements.
The Limit annunciator will go on whenever a result is out of limits. As soon as a
result is within limits, the Limit annunciator will go off.
Limits Testing Example 3—Use Analog Graph Display While
Adjusting/Aligning Input Signal
1 For demonstration purposes, connect a 10 MHz signal to channel 1, and press
Freq Ch 1 key.
2Using the Uppr & Lower key, set the upper limit to 10.2 MHz and the lower limit
to 9.8 MHz. (Refer to the section in Chapter 1 titled “To Set Limits of
Measurements” if needed.)
3 Press Limit Modes key until SHOW: NUMBER is displayed.
4 Press any one of the arrow keys to toggle to the next state.
SHOW: GRAPH is displayed.