To Measure Positive/Negative Volt Peaks
Press Other Meas key until VOLT PEAKS 1 is displayed.
VOLT PEAKS 1 is momentarily displayed, the Ch1 annunciator lights, and the
Counter is ready to measure minimum and maximum voltage peaks of signals applied
to CHANNEL 1.
NOTE In Volt Peaks function, the Counter measures the input signal after the signal has been
conditioned by the input settings (impedance, coupling, attenuation, and filter). One
reminder of this is the (AC COUPLED) message, which is displayed when Voltage
Peaks is selected and the measurement channel is ac coupled.
The result is corrected for X10 attenuation.
NOTE Arming is not used for Voltage Peaks measurements. Also, Limits, Math, and Display
Digits are not available for Voltage Peaks measurements.