Chapter 2 Operating Your Frequency Counter
Using the Utility Menu
Operating Guide 2-59
2
Setting the Baud Rate
You can select one of five baud rates for RS-232 operation. The rate is set to 9600
baud when the Counter is shipped from the factory.
1 Press Recall (Utility) key until BAUD: is displayed.
2 Press any one of the arrow keys until the desired baud rate is displayed.
The baud rate is stored in non-volatile memory, and does not change when power has
been off or after a remote interface reset.
Setting the Parity
You can select the parity for RS-232 operation. The Counter is configured for parity
off when shipped from the factory.
1 Press Recall (Utility) key until PARITY: is displayed.
2 Press any one of the arrow keys until the desired parity is displayed.
The parity choice is stored in non-volatile memory, and does not change when power
has been off or after a remote interface reset.
Setting the Software Pace
You can select the software pace for RS-232 operation.
1 Press Recall (Utility) key until SW PACE: is displayed.
2 Press any one of the arrow keys until the desired software pacing is displayed.
The software pacing choice is stored in non-volatile memory, and does not change
when power has been off or after a remote interface reset.