Chapter 2 Operating Your Frequency Counter
Using the Save and Recall Menus
Operating Guide 2-49
2
To Use the Save Function
The following step-by-step procedure describes how to SAVE a simple measurement
setup to quickly get you familiar with the Save function.
You will SAVE the following in register 1 (SAVE: 1):
the Period measurement function,
the input conditioning parameters:
auto trigger to off
trigger level to 1.5V
input impedance to 50
input coupling to DC
1Cycle POWER key to preset the Counter setup.
The Counter is now ready to measure the frequency of a signal applied to
CHANNEL 1 input as indicated by the Freq and Ch1 annunciators.
2 Press Other Meas key until PERIOD 1 is displayed.
The Counter is now ready to measure the period of a signal applied to CHANNEL 1
input as indicated by the Period and Ch1 annunciators.
3 Press Trigger/Sensitivity key.
AUTO TRG: ON is displayed.
4 Press any one of the arrow keys.
AUTO TRG: OFF is displayed.
5 Press Trigger/Sensitivity key, or Enter key.
LEVEL: 0.000v is displayed.
6 Using the appropriate arrow keys, set the trigger level to 1.5V.