Chapter 2 Operating Your Frequency Counter
Using the LIMITS Menu Keys
2-36 Operating Guide
2
8 To change the Counter back to displaying numbers, press Limit Modes key until
SHOW: GRAPH is displayed.
9 Press any one of the arrow keys to toggle to SHOW: NUMBER.
10 Press Run key to display measurements as numbers.
Limits Testing Example 4—Selecting Filtering Conditions of
Stats Computation
Lets assume you have set the upper and lower limits to reasonable values as in the
previous procedure.
NOTE Since the Limit Testing and Stats functions are independent, LIM TEST: doesn’t have
to be ON in order to filter measurements for statistics.
1 Press Stats key until USE: ALL MEAS is displayed.
2 Press any one of the arrow keys to display USE: IN LIMIT.
3 Press Run key to start the measurement.
The Counter will compute statistics using only measurements that are within the user-
entered limits.
4 Press Stats key until USE: IN LIMIT is displayed.
5 Press any one of the arrow keys to display USE: ALL MEAS.
The Counter will compute statistics using all measurements, whether they fall in or
out of the user-entered limits.