Chapter 2 Operating Your Frequency Counter
Front Panel Display Messages
2-66 Operating Guide
2
1
If this condition recurs, your Counter may need servicing.
NO REGISTERS There are no registers which can be recalled and you selected the Recall menu.
NO STAT MENU You currently have the VOLT PEAKS function selected and requested the Stats menu.
NO TRIG MENU You currently have the VOLT PEAKS function selected and requested the
Trigger/Sensitivity menu.
NOT ENTERED The Enter key was NOT pressed after you changed a numeric parameter. The parameter
(for example, SCAL: nnnnn) was returned to its previous value and must be re-entered.
Press the Enter key to save changes made to numeric parameters.
This message will also be displayed when an event (for example, TEST: ALL or SET
REF: YES) or help (MATH HELP: YES) is selected, but not invoked with Enter key.
OFFS 1 FAIL You have invoked the OFFS choice from the Calibration menu item CAL:, and the
calibration failed.
OFFS 1 PASS You have invoked the OFFS choice from the Calibration menu item CAL:, and the
calibration passed.
1
RECALL FAIL You have requested the Counter to perform a recall (from the Recall menu) and a
hardware failure has resulted.
1
SAVE FAIL You have requested the Counter to perform a save (from the Save & Print menu) and a
hardware failure has resulted.
TB CAL FAIL You have invoked the TIMEBAS choice from the Calibration menu item CAL:, and the
calibration failed.
TB CAL PASS You have invoked the TIMEBAS choice from the Calibration menu item CAL:, and the
calibration passed.
Table 2-4. Menu Messages (continued)
Display Message Description