Chapter 3 Specifications
Introduction
3-14 Operating Guide
Frequency Ratio:
Measurement is specified over the full signal range of each input.
Results Range: 10
–10
to 10
11
‘Auto Gate Time: 100 ms (or sufficient cycles on Channel 1 to make a valid measurement, whichever is longer)
LSD:
RMS Resolution:
To minimize relative phase measurement error, connect the higher frequency signal to Channel 1 when possible.
Peak Volts
Measurement is specified on Channels 1 for DC signals; or for AC signals of frequencies between 100 Hz and 30 MHz with peak-to-peak
amplitude greater than 100 mV. (The measurement will continue to operate up to 225 MHz, though results are for indication only.)
Results Range: –5.1 V to +5.1 V
Resolution: 10 mV
Systematic Uncertainty for AC signals: 25 mV + 10% of V
for DC signals: 25 mV + 2% of V
Use of the input attenuator multiplies all voltage specifications (input range, results range, resolution and systematic uncertainty) by a
nominal factor of 10. For example with AC signals, the Systematic Uncertainty becomes: 250 mV + 10% of V.
Measurement Specifications (Continued)
h1
Ch2
Ch2
Ch
R
atio
1
2
:
1
Ch2 Freq Gate Tim
e
×
-------------------------------------------------------
-----
R
atio
2
1
:
Ch2 Freq
Ch1 Freq()
2
Gate Tim
e
×
--------------------------------------------------------------
-----
R
atio
1
2
:
2 1 Ch1 Freq Ch2 Trigger Erro
r
×()
2
+×
Ch2 Freq Gate Time×
---------------------------------------------------------------------------------------------------------
------
R
atio
2
1
:
2 Ch2 Freq 1 Ch1 Freq Ch2 Trigger Erro
r
×()
2
+××
Ch1 Freq()
2
Gate Time×
-----------------------------------------------------------------------------------------------------------------------------------------
------