Chapter 2 Operating Your Frequency Counter
Using the Measurement Control Keys (Run and Stop/Single)
2-8 Operating Guide
2
7 Press Stop/Single key.
One set of N frequency measurements is taken with each press of the Stop/Single key.
One set of statistics is computed with each press of the Stop/Single key.
8 Press Stats key until SHOW menu item appears.
9 Press any one of the arrow keys to select a statistical result to display, and press
Freq Ch 1 key.
The Counter is now displaying one of the statistical results.
10 Press Stop/Single key.
One set of N frequency measurements is taken with each press of the Stop/Single key.
At the completion of the Nth measurement, the Counter displays the statistical result
selected in step 9.
11 Press Stats key until ON SINGLE menu item appears, then press any one of the
arrow keys to set ON SINGLE menu item to 1.
The Counter is set to take one measurement on each Stop/Single keypress.
Disabling the statistics (with the STATS: OFF menu item) is another way to cause
the Stop/Single key to initiate one measurement at a time.