EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7200

Agilent Technologies 7200 Troubleshooting And Maintenance Guide

Agilent Technologies 7200
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
36 Troubleshooting and Maintenance Manual
2 General Troubleshooting
Pressure Symptoms
This section describes unusual pressure readings and their possible causes. At
typical column flow rates (0.5 to 2.0 mL/minute), the foreline pressure will be
approximately 16 to 18 mTorr. The Quad pressure with collision cell gas on or
off will be approximately 1 × 10
-4
to 2 × 10
-4
Torr. These pressures can vary
widely from instrument to instrument so it is very important that you are
familiar with the pressures that are typical for your instrument at given
carrier and collision gas flows.
Foreline pressure is too high
If the pressure you observe for a given column flow has increased over time,
check the following:
Column (carrier gas) flow is too high
Collision cell gas flow is too high
Air leak (usually the side plate is not pushed in or vent valve is open)
Foreline pump oil level is low or oil is contaminated
Foreline hose is constricted
Foreline pump is not working correctly

Table of Contents

Other manuals for Agilent Technologies 7200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7200 and is the answer not in the manual?

Agilent Technologies 7200 Specifications

General IconGeneral
BrandAgilent Technologies
Model7200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals