General Maintenance 4
Troubleshooting and Maintenance Manual 83
To Move or Store the MS
Materials needed
Ferrule, blank (5181-3308)
Interface column nut (05988-20066)
2 Wrenches, open-end, 1/4-inch × 5/16-inch (8710-0510)
Procedure
1 Cool down the GC and MS. See the Agilent 7200 Accurate-Mass Q-TOF
GC/MS System Operation Manual.
2 Vent the MS. See the Agilent 7200 Accurate-Mass Q-TOF GC/MS System
Operation Manual.
WARNING
Make sure the GC/MS interface and the analyzer zones are cool (below 100 °C)
before you vent the MS. A temperature of 100 °C is hot enough to burn skin; always
wear cloth gloves when handling analyzer parts.
WARNING
The use of hydrogen gas is specifically prohibited with this product.
WARNING
When the MS is vented, do not put the MassHunter Workstation software into
Instrument Control view. Doing so will turn on the interface heater.
CAUTION
Be sure the GC oven and the GC/MS interface are cool before turning off the carrier
gas flow.
CAUTION
Never vent the MS by allowing air in through either end of the foreline hose. Always
use the automated procedure in MassHunter Data Acquisition to vent the MS.
Do not exceed the maximum recommended total gas flow.

Table of Contents

Other manuals for Agilent Technologies 7200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7200 and is the answer not in the manual?

Agilent Technologies 7200 Specifications

General IconGeneral
BrandAgilent Technologies
Model7200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals