General Troubleshooting 2
Troubleshooting and Maintenance Manual 29
Chromatographic Symptoms
These are symptoms you may observe in the chromatograms generated by data 
acquisition. In general, these symptoms do not prevent you from operating 
your GC/MS system. They indicate, however, that the data you are acquiring 
may not be the best data obtainable. These symptoms can be caused by 
instrument malfunctions but are more likely caused by incorrect 
chromatographic technique.
Two of the symptoms, Low sensitivity and Poor repeatability also apply to 
mass spectral data.
No peaks
If an analysis shows no chromatographic peaks, only a flat baseline or minor 
noise, run the automated tune program. If the MS passes tune, the problem is 
most likely related to the GC. If the MS does not pass tune, the problem is most 
likely in the MS.
Passes tune
• Incorrect sample concentration
• No analytes present
• Syringe missing from the ALS or not installed correctly
• Injection accidentally made in split mode instead of splitless mode
• Empty or almost empty sample vial
• Dirty GC inlet
• Leaking GC inlet*
• Loose column nut at the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC 
from operating.
Does not pass tune
• Calibration vial is empty
• Excessive foreline or analyzer chamber pressure
• Ver y dirty ion source