General Troubleshooting 2
Troubleshooting and Maintenance Manual 39
MS electronics are not working correctly
* This will cause an error message.
GC/MS interface will not heat up
Incorrect setpoint(s)
Setpoint entered in wrong heated zone
GC/MS interface has not had enough time to reach temperature setpoint
GC is off
GC experienced a fault and needs to be reset*
GC/MS interface heater/sensor cable is not connected*
GC/MS heater failed (burned out)*
GC/MS sensor failed*
GC electronics are not working correctly*
* This will cause a GC error message. GC error messages are described in
the documentation supplied with your GC.