EasyManuals Logo

Agilent Technologies 7200 Troubleshooting And Maintenance Guide

Agilent Technologies 7200
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
General Troubleshooting 2
Troubleshooting and Maintenance Manual 47
Table 2 Common contaminants
Ions (m/z) Compound Possible source
18, 28, 32, 44 or 14, 16 H
2
O, N
2
, O
2
, CO
2
or N, O Residual air and water, air
leaks, outgassing from Vespel
ferrules
31, 51, 69, 100, 119, 131,
169, 181, 214, 219, 264, 376,
414, 426, 464, 502, 576, 614
PFTBA and related ions PFTBA (tuning compound)
31 Methanol Cleaning solvent
43, 58 Acetone Cleaning solvent
78 Benzene Cleaning solvent
91, 92 Toluene or xylene Cleaning solvent
105, 106 Xylene Cleaning solvent
151, 153 Trichloroethane Cleaning solvent
69 Foreline pump oil or PFTBA Foreline pump oil vapor or
calibration valve leak
73, 147, 207, 221, 281, 295, 355,
429
Dimethylpolysiloxane Septum bleed or methyl
silicone column bleed
149 Plasticizer (phthalates) Vacuum seals (O-rings)
damaged by high
temperatures, vinyl gloves
Peaks spaced 14 m/z apart Hydrocarbons Fingerprints, foreline pump oil
50, 69, 76, 100, 119, 126, 171,
221, 271, 366, 416, 435
PFET and related ions PFET (IRM calibrant)

Table of Contents

Other manuals for Agilent Technologies 7200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7200 and is the answer not in the manual?

Agilent Technologies 7200 Specifications

General IconGeneral
BrandAgilent Technologies
Model7200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals