Installation 1
Installation and First Startup 15
Place the Headspace Sampler on the Bench
The HS requires a benchtop that can support its weight plus
that of other equipment you will use with it. The area must
be free of overhanging obstructions that might interfere with
cooling and limit access to the top of the instrument.
1 Remove the HS from its shipping box.
Be careful when removing the packing foam from
around the instrument. The foam also secures the
transfer line.
2 Place the HS on the benchtop. Make sure gas and power
supplies are accessible. Place other required equipment
near the HS. Install only on the right side of the GC.
3 Remove caps and packing.
Protective caps cover gas input and vent fittings on the
back of the unit.
Remove any packing tape that secures the transfer line
against the instrument.
For the 12 vial model, lift the tray cover and remove any
packing material from the tray.
WARNING
The instrument is heavy. Use extreme caution when handling the
instrument. A two person lift is recommended. Failure to perform a
two person lift may result in personal injury.

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals