Installation 1
Installation and First Startup 51
Configure the Headspace Sampler
Next, enter details that describe the gas types, vial sizes,
sample loop size, and other important information.
1 Press [Config].
2 Use  to scroll to Vial Gas type and press [Mode/Type].
3 Use  to select the vial pressurization gas type from the
list, then press [Enter].
4 Scroll to Loop Volume (mL). Press [Mode/Type]. Select 1.0 as
the sample loop volume, in mL, then press [Enter]. (All
headspace samplers ship with a 1 mL sample loop
installed.)
5 If you ordered the optional G4562A Carrier Gas EPC
Module Accessory, scroll to Carrier. Press [Mode/Type].
Select the correct control mode for checkout:
If the HS will provide all carrier gas to the GC, select
GC Control. (For example, you did not plumb carrier gas
to the GC.)
If the GC will provide its own carrier gas flow, select
GC + HS Control. In this case, the HS will provide enough
additional carrier gas flow during injection to push the
checkout sample into the GC inlet.
6 Scroll to APG polarity. If needed, press [Mode/Type] to
change.
For Agilent GCs, use Active high.
For non- Agilent GCs, refer to the GC documentation.

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals