EasyManuals Logo

Agilent Technologies 7697A Installation And Startup Guide

Agilent Technologies 7697A
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
62 Installation and First Startup
2 First Startup
Run the Checkout Sample
After the HS has been Ready for 25–30 minutes, it has been
purged and is sufficiently stable to run the checkout sample.
1 Load the checkout sample into the HS.
For the 111 vial model:
a Press [Tray Park/Carousel Advance] to “park” the tray
(move the gantry to a rest position for easy access to
the vial racks).
b Place the capped sample vial in position 1.
Figure 19 Tray vial position 1
c Press [Tray Park/Carousel Advance] to prepare the tray for
use.
For the 12 vial model:
a Open the tray cover. A label to the left of each vial
position denotes the number of that position. See
Figure 19 and Figure 20.
b If vial position 1 is not available, press [Tray
Park/Carousel Advance] to rotate the tray.
c Place the capped sample vial in position 1.
1
2
3
4
5
7
1
2
1
1

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals