62 Installation and First Startup
2 First Startup
Run the Checkout Sample
After the HS has been Ready for 25–30 minutes, it has been
purged and is sufficiently stable to run the checkout sample.
1 Load the checkout sample into the HS.
For the 111 vial model:
a Press [Tray Park/Carousel Advance] to “park” the tray
(move the gantry to a rest position for easy access to
the vial racks).
b Place the capped sample vial in position 1.
Figure 19 Tray vial position 1
c Press [Tray Park/Carousel Advance] to prepare the tray for
use.
For the 12 vial model:
a Open the tray cover. A label to the left of each vial
position denotes the number of that position. See
Figure 19 and Figure 20.
b If vial position 1 is not available, press [Tray
Park/Carousel Advance] to rotate the tray.
c Place the capped sample vial in position 1.
1
2
3
4
5
7
1
2
1
1