Making Swagelok Connections A
Installation and First Startup 73
Using a Swagelok Tee
To supply gas from a single source to more than one input,
use a Swagelok Tee.
Materials needed:
1/8- inch preconditioned copper tubing
Tubing cutter
1/8- inch Swagelok nuts and front and back ferrules
1/8- inch Swagelok Tee
Two 7/16- inch wrenches
1/8- inch Swagelok cap (optional)
1 Cut the tubing where you want to install the Tee. Connect
the tubing and Tee with a Swagelok fitting. See Figure 31.
Figure 31 Swagelok tee
2 Measure the distance from the Tee to the instrument
fittings. Attach copper tubing to the open Tee ends with
Swagelok fittings.
NOTE
Do not combine valve actuator air with flame ionization air. The valve
action will cause major upsets in the detector signal.
Cap
Tee fitting
Nut and
ferrules