First Startup 2
Installation and First Startup 57
Make the First Connection
If using an Agilent data system, perform this step. If
operating the HS in standalone mode, skip this section.
1 Open the online session for the GC.
2 As the GC session opens, you will be prompted to
configure the instrument. Select Yes. The instrument
configuration utility opens.
If not prompted, open the utility from the Instrument
menu.
3 In the left pane, double- click on the 7697A instrument.
The instrument icon moves to the right hand list.
4 Click Configure. In the dialog that opens, enter the
following:
• In the Connect Info group, enter the Instrument Name,
then enter the instrument IP address or host name.
• Enter your license key in the Instrument License field.
You will see the Version Information for the instrument
appear (instrument serial number, firmware version,
software version, and so forth).
• Enter any Notes as desired.