18 Installation and First Startup
1 Installation
Install the configuration plug
Install the configuration plug as follows:
Table 6 Power consumption
Product Line voltage (VAC) Frequency (Hz) Maximum
continuous
power
consumption
(VA)
Current rating
(amps)
Power outlet
current rating
111 Vial model 100*/120
single phase
(–10% / +10%)
50/60 850 6.2 15 Amp
111 Vial model 200/220/230/240
single/split phase
(–10% / +10%)
50/60 850 3.8/3.4/3.3/3.1 10 Amp
12 Vial model 100*/120
single phase
(–10% / +10%)
50/60 850 6.2 15 Amp
12 Vial model 200/220/230/240
single/split phase
(–10% / +10%)
50/60 850 3.8/3.4/3.3/3.1 10 Amp
* Users of 100 V power (Japan) receive the 120 V US power option with a 100 V/120 V step up transformer.
WARNING
Only an Agilent-trained technician should perform this procedure.
CAUTION
Components can be damaged by static electricity: be sure to wear
an ESD strap grounded to the GC chassis while performing this
procedure.

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals