EasyManua.ls Logo

Agilent Technologies 7697A - Install the Configuration Plug

Agilent Technologies 7697A
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18 Installation and First Startup
1 Installation
Install the configuration plug
Install the configuration plug as follows:
Table 6 Power consumption
Product Line voltage (VAC) Frequency (Hz) Maximum
continuous
power
consumption
(VA)
Current rating
(amps)
Power outlet
current rating
111 Vial model 100*/120
single phase
(–10% / +10%)
50/60 850 6.2 15 Amp
111 Vial model 200/220/230/240
single/split phase
(–10% / +10%)
50/60 850 3.8/3.4/3.3/3.1 10 Amp
12 Vial model 100*/120
single phase
(–10% / +10%)
50/60 850 6.2 15 Amp
12 Vial model 200/220/230/240
single/split phase
(–10% / +10%)
50/60 850 3.8/3.4/3.3/3.1 10 Amp
* Users of 100 V power (Japan) receive the 120 V US power option with a 100 V/120 V step up transformer.
WARNING
Only an Agilent-trained technician should perform this procedure.
CAUTION
Components can be damaged by static electricity: be sure to wear
an ESD strap grounded to the GC chassis while performing this
procedure.

Other manuals for Agilent Technologies 7697A

Related product manuals