EasyManua.ls Logo

Agilent Technologies 7697A - Page 59

Agilent Technologies 7697A
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
First Startup 2
Installation and First Startup 59
Total flow 87 mL/min
Gas saver Off
Column
Mode Constant flow
Initial flow 4.0 mL/min
Initial pressure 134.4 kPa (19.5 psi)
Average velocity 60 cm/s
Detector parameters–FID
Temperature 300 °C
Hydrogen (fuel) flow 30 mL/min
Air flow 400 mL/min
Makeup mode Constant makeup
Makeup flow 25 mL/min
Detector parameters–NPD
Temperature 300 °C
Hydrogen (fuel) flow 2.0 mL/min
Air flow 60 mL/min
Makeup mode Constant makeup
Makeup flow 8 mL/min
Detector parameters–µECD
Temperature 250 °C
Makeup mode Constant makeup
Makeup flow 30 mL/min
MS or MSD
Solvent delay 3 min
Run time 6.5 min
Headspace event time for GC cycle
time
7 min
Scan range 45 to 300
Method used Atune file
Inlet pressure 82.8 kPa (12 psi)
Column flow 1.1 mL/min
Table 11 GC checkout method (continued)

Other manuals for Agilent Technologies 7697A

Related product manuals