6-63
Calibrating for Increased Measurement Accuracy
Calibrating Using Electronic Calibration (ECal)
Figure 6-22 Manual Thru Setup
5. After you connect the manual thru, press to complete the manual
thru portion of the ECal.
6. If you are calibrating using two ECal modules, a prompt is displayed directing you to
remove the first module and connect the second module. Follow this prompt as shown in
Figure 6-23.
Figure 6-23 Connect the Second ECal Module
CONTINUE ECal