7-3
Options and Accessories
Analyzer Options Available
Analyzer Options Available
Option 1D5, High Stability Frequency Reference
Option 1D5 offers ±0.05 ppm temperature stability from 0 to 55 °C (referenced to 25 °C),
and aging rate of ±0.5 ppm per year (typical).
Option 002, Harmonic Mode
Provides measurement of second or third harmonics of the test device's fundamental
output signal. Frequency and power sweep are supported in this mode. Harmonic
frequencies can be measured up to the maximum frequency of the receiver. However, the
fundamental frequency may not be lower than 16 MHz.
Option 004, Source Attenuator (ET Only)
This option adds a 55 dB step attenuator that extends the output power range by allowing
lower power levels.
Option 006, 6 GHz Operation
Option 006 extends the maximum source and receiver frequency of the analyzer to 6 GHz.
Option 010, Time Domain
This option displays the time domain response of a network by computing the inverse
Fourier transform of the frequency domain response. It shows the response of a test device
as a function of time or distance. Displaying the reflection coefficient of a network versus
time determines the magnitude and location of each discontinuity. Displaying the
transmission coefficient of a network versus time determines the characteristics of
individual transmission paths. Time domain operation retains all accuracy inherent with
the correction that is active in the frequency domain. The time domain capability is useful
for the design and characterization of such devices as SAW filters, SAW delay lines, RF
cables, and RF antennas.
Option 011, Delete Built-In Test Set (ES Only)
Option 011 allows front panel access to the source output and the R, A, and B sampler
inputs. The transfer switch, couplers, and bias tees have been removed. Therefore,
external accessories are required to make most measurements.
Option 014, Configurable Test Set (ES Only)
This option adds SMA connectors on the front panel for access to various source output
and receiver input signal paths, including direct access to receivers. When jumpers are in
place, the analyzer performs like a standard 8753ES network analyzer except that the
maximum source output power is reduced by 2 dB. Not compatible with Option 011 or 075.

Table of Contents

Other manuals for Agilent Technologies 8720ES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720ES and is the answer not in the manual?

Agilent Technologies 8720ES Specifications

General IconGeneral
BrandAgilent Technologies
Model8720ES
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals