Troubleshooting the Analyzer
To troubleshoot memory battery failures
To troubleshoot memory battery failures
Use this test when battery-backed-up memory is suspected of failing.
This test separates A42 Memory assembly failures from memory battery
failures.
1 Check battery-backed-up memory.
a Set the power switch to on ( l ).
b Press the following keys:
[System Utility]
[
more]
[
time/date setup]
[
date]
010193
[enter
]
c Set the power switch to off ( O ), then to on ( l ).
d Press the following keys:
[System Utility]
[
more]
[
time/date setup]
e If the date is 01-01-93, the battery-backed-up memory is probably
functioning correctly. Go to page 1-32, ‘’To perform self tests,’’ to
continue troubleshooting.
2 Check the battery voltage.
a Set the power switch to off ( O ).
b Remove the screws on each side and across the bottom of cover
number 6.
c Pull out the assemblies attached to cover number 6 until the
battery on the Memory assembly is visible.
d Measure the voltage between the top of the battery and A42 TP18.
e If the voltage is 3 ±1 V, the A42 Memory assembly is probably
faulty.
f If the voltage is incorrect, replace the battery (A42 B100).

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals