2-34
Adjusting the Analyzer
To adjust oven frequency
To adjust oven frequency
This procedure adjusts the optional A85 Oven assembly’s frequency. The
Agilent 89410A must be on for at least 48 hours before this adjustment is
made.
Equipment Required:
Oscilloscope
Frequency standard
BNC cable
Flat-edge adjustment tool, Agilent part number 8710-1928
WARNING: This procedure is performed with protective covers removed and
power applied. Energy available at may points can, if contacted, result in personal
injury.
Even with power removed, there can be sufficient stored energy in some circuits to
cause personal injury. These voltages will discharge to a relatively safe level
approximately five minutes after the power cord is disconnected.
1 Remove the power supply shield.
a Set the power switch to off ( O ).
b Remove the power supply shield. The power supply shield is
located behind the display.
c Set the power switch to on ( l ).

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals