2-25
Adjusting the Analyzer
To adjust 10 MHz low pass filter
To adjust 10 MHz low pass filter
This procedure adjusts the A35 Analog Source assembly’s reconstruction
filter. The reconstruction filter is a 10 MHz low pass filter.
Equipment Required:
Network analyzer
Extender board, Agilent part number 89410-B1001
(2) BNC cables
(2) N(m)-to-BNC(f) adapters
BNC(f)-to-test clips
Flat-edge adjustment tool, Agilent part number 8710-1928
1 Place the A35 Analog Source assembly on an extender board.
a Set the power switch to off ( O ).
b Remove screws holding the A35 Analog Source assembly in the
card nest. Do not remove the screws in the center of the cover.
c Place the assembly on an extender board.
d Set the power switch to on ( l ).
2 Set up the Agilent 89410A.
a Press the following keys:
[ System Utility ]
[ more ]
[
diagnostics ]
[
service functions ]
1125
[ enter ]
[
adjustments ]

Table of Contents