2-26
Adjusting the Analyzer
To adjust 10 MHz low pass filter
[ analog source ]
[
setup 1 ]
3 Set up the network analyzer.
a Set the network analyzer as follows:
b Connect the network analyzer’s reference channel to the
Agilent 89410A’s source using an N(m)-to-BNC(f) adapter and
BNC cable.
c Connect the network analyzer’s source to A35 TP14 (AMPIN) and
TP13 (Ground) using an N(m)-to-BNC(f) adapter, BNC cable, and
BNC(f)-to-test clips adapter.
Receiver
Attenuation
Resolution BW
Display format
reference channel
20 dB
100 Hz
log magnitude
Source
Sweep Mode
Sweep Type
Sweep Time
Amplitude
Start Frequency
Stop Frequency
continuous
linear frequency sweep
3 s
-40 dBm
10 MHz
30 MHz