EasyManua.ls Logo

Agilent Technologies 89410A - To Troubleshoot Source and Calibrator out Failures

Agilent Technologies 89410A
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1-62
Troubleshooting the Analyzer
To troubleshoot source and calibrator out failures
To troubleshoot source and calibrator out failures
Use this test when the analog source or calibrator is suspected of failing
and all self tests listed before the ADC in the Self-Test Troubleshooting
Guide on page 1-32 passed.
1 Check the sine wave output.
a Set the power switch to on ( l ).
b Using a BNC cable, connect an oscilloscope to the SOURCE
output.
c Set the oscilloscope as follows:
CH1 V/div 400 mV/div
Input Impedance 50
CH1 Coupling dc
Probe Atten 1
Display Mode Real Time
Time/div 200 ns/div
Trigger Trg’d Sweep
Trig Src Chan1
Trigger Level 0 V
d Press the following keys:
[System Utility]
[
auto cal off]
[more cal setup
]
[
auto zero cal off]
[Source
]
[
source on]
[level
]
1
[
V]
e The oscilloscope should display a 1 MHz, 2 Vp-p sine wave with no
dc offset. If the signal is incorrect, go to Step 7.
This is only a quick check of the source sine wave. If the source is
suspected of failing at a specific frequency or amplitude, check

Table of Contents

Other manuals for Agilent Technologies 89410A

Related product manuals