EasyManuals Logo

Agilent Technologies 89410A Service Guide

Agilent Technologies 89410A
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #145 background imageLoading...
Page #145 background image
2-29
Adjusting the Analyzer
To adjust auto-range detect level
To adjust auto-range detect level
This procedure adjusts the A36 Trigger assembly’s ±12 Vdc regulators.
The regulators are used to set the over-range and half-range detect levels
in the auto-range circuits.
Equipment Required:
Digital multimeter
Extender board, Agilent part number 89410-B1002
Flat-edge adjustment tool, Agilent part number 8710-1928
Test clips-to-dual banana plug cable
1 Place the A36 Trigger assembly on an extender board.
a Set the power switch to off ( O ).
b Remove screws holding the A36 Trigger assembly in the card nest.
c Place the assembly on an extender board.
d Set the power switch to on ( l ).
2 Adjust A36 R113.
a Set the digital voltmeter to Vdc.
b Connect the digital voltmeter to A36 TP3 and TP1 (ground).
c Adjust A36 R113 for a voltmeter reading of 1.259 ±0.003 Vdc using
the flat-edge adjustment tool.

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals