EasyManua.ls Logo

Agilent Technologies 89410A - To Adjust Input Capacitance

Agilent Technologies 89410A
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-13
Adjusting the Analyzer
To adjust input capacitance
To adjust input capacitance
This procedure adjusts the A10 Analog Input assembly’s input
capacitance for a constant value over all input ranges and attenuator
settings.
Perform the previous adjustment before doing this adjustment.
Equipment Required:
Network analyzer
Extender board, Agilent part number 89410-B1001
BNC cable
SMB(f)-to-BNC(m) cable
50
feedthrough termination
10 k series resistor
(2) N(m)-to-BNC(f) adapters
Flat-edge adjustment tool, Agilent part number 8710-1928
1 Set up the Agilent 89410A.
a Disconnect the network analyzer’s source and receiver from the
Agilent 89410A.
b Press the following keys:
[ Return ]
[ input capacitance ]
[
setup 1 ]

Table of Contents

Other manuals for Agilent Technologies 89410A

Related product manuals