2-13
Adjusting the Analyzer
To adjust input capacitance
To adjust input capacitance
This procedure adjusts the A10 Analog Input assembly’s input
capacitance for a constant value over all input ranges and attenuator
settings.
Perform the previous adjustment before doing this adjustment.
Equipment Required:
Network analyzer
Extender board, Agilent part number 89410-B1001
BNC cable
SMB(f)-to-BNC(m) cable
50
feedthrough termination
10 k series resistor
(2) N(m)-to-BNC(f) adapters
Flat-edge adjustment tool, Agilent part number 8710-1928
1 Set up the Agilent 89410A.
a Disconnect the network analyzer’s source and receiver from the
Agilent 89410A.
b Press the following keys:
[ Return ]
[ input capacitance ]
[
setup 1 ]