Page #174 background imageLoading...
Page #174 background image
3-20
Replacing Assemblies
To remove rear panel
To remove rear panel
1
Using a T-15 torx driver, remove the screw from the back of
the bottom cover. Slide the cover off.
2
Using a T-15 torx driver, remove the screw from the back of
the top cover. Slide the cover off.
3
Using a 2 point pozidriv, remove the screw from both end
caps of both strap handles. Remove the strap handles and
side covers.
4
Using a T-10 torx driver, remove the screws from the
assembly covers.