EasyManua.ls Logo

Agilent Technologies 89410A - Page 144

Agilent Technologies 89410A
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-28
Adjusting the Analyzer
To adjust 10 MHz low pass filter
5 Adjust A35 L3 and A35 L4.
a Press the
[ setup 2 ] softkey.
b Set the network analyzer’s reference level to 10 dB.
c Set the scale to 10 dB/div.
d Adjust A35 L3 for zero transmission at 15.8 MHz using the
flat-edge adjustment tool.
e Adjust A35 L4 for zero transmission at 22.8 MHz using the
flat-edge adjustment tool. L3 and L4 correctly adjusted
6 Reinstall the A35 Analog Source assembly.
a Set the power switch to off ( O ).
b Place the assembly in the card nest.
If only the source assembly adjustments are being done, go to
page 2-37 to adjust the calibrator. The calibrator adjustment must
be done after the channel 1 A10 Analog Input assembly or A35
Analog Source assembly is adjusted.
This completes the A35 Analog Source assembly adjustments.

Table of Contents

Other manuals for Agilent Technologies 89410A

Related product manuals