EasyManua.ls Logo

Agilent Technologies 89410A - 13 Using Wideband CDMA (Options B73, B79, and 080)

Agilent Technologies 89410A
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-35
Adjusting the Analyzer
To adjust oven frequency
2 Set up the oscilloscope.
a Connect the frequency standard’s output to the oscilloscope’s
external trigger. Check that the frequency standard is properly
terminated.
b Set the oscilloscope as follows:
c Connect the rear panel OVEN REF OUT connector to the
oscilloscope’s channel 1 input.
3 Adjust A85 HR1.
a Adjust A85 R12 to the center of its adjustment range.
b Remove the oven adjustment screw.
c Adjust the A85 HR1, FREQ ADJ, for a stable (not moving) display
on the oscilloscope.
4 Adjust A85 R12.
a Change the oscilloscope timebase to 10 ns per division.
b Adjust A85 R12 for a stable display on the oscilloscope.
Channel 1
Volts/Div
Offset
Coupling
100 mV/div
0 V
1 M
ac
Timebase
Time/Div
Sweep
50 ns/div
Triggered
Trigger
Level
Slope
Mode
Source
0 V
Positive
Edge
External Trigger
Display
Mode
Screen
Real Time
Single

Table of Contents

Other manuals for Agilent Technologies 89410A

Related product manuals