2-36
Adjusting the Analyzer
To adjust oven frequency
5 Replace the oven screw and power supply shield.
a Set the power switch to off ( O ).
b Replace the oven adjustment screw.
c Replace the power supply shield.
This completes the A85 Oven assembly adjustments.