2-38
Adjusting the Analyzer
To adjust calibrator
2 Set up the synthesizer.
a Set the synthesizer as follows:
b Connect the synthesizer’s external reference to the
Agilent 89410A’s EXT REF IN. See the following illustration for
setup.
c Connect the synthesizer’s output to the milliwatt power meter.
d Adjust the synthesizers amplitude to read 0 dBm on the milliwatt
power meter.
3 Determine and store the exact amplitude of the calibrator signal.
a Connect the synthesizer’s output to Agilent 89410A’s channel 1.
b Press the
[ single cal ] softkey.
c Wait for the program to finish.
The program takes about 2 minutes to complete.
The exact amplitude of the calibrator signal is now stored in Monitor
Flash ROM.
Frequency
Amplitude
Function
3.2 MHz
0dBm
sine wave