EasyManua.ls Logo

Agilent Technologies 89410A - To set up transient analysis

Agilent Technologies 89410A
184 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1-24
Troubleshooting the Analyzer
To troubleshoot the power supply
9 Isolate assembly loading power supply.
a Set the power switch to off ( O ) and disconnect the power cord
from the rear panel.
b Wait five minutes to allow time for the power supply capacitors to
discharge.
c Disconnect A95 P360 pins 2 and 3.
d Reconnect all the cables to the A95 Main Power Supply assembly.
e Reinstall the A95 Main Power Supply assembly in the analyzer.
f Replace the black power supply shield and bottom cover.
g Repeat the following steps for each assembly suspected of loading
the power supply.
Use the Power Supply Voltage Distribution table on page 6- and
the A90/A91 Motherboard Voltages table on page 6- to help isolate
the failing assembly.
h Set the power switch to off ( O ).
i Remove the suspected assembly or the power cable to the
assembly.
j Set the power switch to on ( l ).
k Check the voltages in the table on page 1-21.

Table of Contents

Other manuals for Agilent Technologies 89410A

Related product manuals