1-33
Troubleshooting the Analyzer
To perform self tests
1 Check sample clocks.
a Remove the top cover.
b Set the power switch to on ( l ).
c Using an oscilloscope and 2 BNC(m)-to-SMB(f) cables, check the
following signals.
d If either signal is incorrect, the A60 Frequency Reference
assembly is probably faulty.
e Reconnect the cables to A60 J3 and A60 J2.
Oscilloscope Setup Parameters Waveform
Connect CH1 to A60 J3
Connect CH2 to A60 J2
Time
Duty Cycle
Time Relationship
CH1 V/div
Input Impedance
CH1 Coupling
Probe Atten
1 V/div
50 Ω
dc
1
Sample Clock
CH2 V/div
Input Impedance
CH2 Coupling
Probe Atten
1 V/div
50 Ω
dc
1
Display Mode
Averaging
Time/div
Tri gger
Trig Src
Repetitive
8
10 ns/div
Trg’d Sweep
Chan1