1-38
Troubleshooting the Analyzer
To perform self tests
3 Step 3. Check for failing functional tests.
a Press the following keys:
[clear test log]
[
Return]
[
functional tests]
[
ALL]
The power-up tests are ran at the beginning of the functional all
self tests. Refer to Step 1 for power-up test failures. All the
power-up tests need to pass (except for calibration) before using
the Self-Test Troubleshooting Guide.
The test log can be printed by connecting a printer to the serial
port or HPIB port. To print the test log to a HPIB printer, press
the following keys:
[
Local/Setup]
[
system controller]
[
peripheral addresses]
[
printer adrs]
printer address
[enter]
[
Plot/Print]
[
output fmt]
[
device defaults]
printer type
[Return]
[
pcl]
[
Return]
[
output to]
[
HPIB]
[
Return]
[
plot item]
[
all display]
[
System Utility]
[
more]
[
diagnostics]
[
test log on]
[Plot/Print
]
[
start plot/print]
b If the analyzer did not finish the tests (analyzer locks up), go to
page 1-48, ‘’To troubleshoot self-test lockup failures.’’