1-47
Troubleshooting the Analyzer
To perform self tests
7 Check the probe power fuse on the A95 Main Power Supply assembly.
a Set the power switch to off ( O ).
b Turn the analyzer upside down.
c Remove the bottom cover.
d Wait five minutes to allow time for the power supply capacitors to
discharge.
e Remove the black power supply shield.
f Measure the resistance across F1.
g If the resistance is 0
Ω, the A92 Probe Power assembly is probably
faulty.
h If the resistance was greater than 0Ω, replace the fuse.