EasyManua.ls Logo

Agilent Technologies Cary 630 FTIR - Page 21

Agilent Technologies Cary 630 FTIR
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting Started
Agilent Cary 630 FTIR Spectrometer User’s Guide 21
Laser Frequency Calibration test
The Laser Frequency Calibration test measures the frequency (X-axis)
accuracy. The test is conducted by measuring a spectrum of a
polystyrene film. Absorbance frequencies from this spectrum are
compared to frequencies set by NIST SRM 1921 polystyrene film. The
software allows administrative-level users to use the test results to set
the laser calibration. Laser calibrations should only be set after
consultation with an Agilent technical support engineer. For this test,
a background is measured first. After the background collection, a
spectrum of polystyrene film is measured, as instructed by the
software. For the Diamond ATR sampling accessory, no sample is
required for a background measurement, but the polystyrene test
sample must be pressed firmly against the diamond sample interface
during the sample collection using the sample press attachment.
Neither the multi-bounce ATR, nor the diffuse reflectance accessory
measurements should be used for the laser frequency calibration test.

Table of Contents

Related product manuals