Maintenance 
60  Agilent Cary 630 FTIR Spectrometer User’s Guide 
5  Position the new window. The locking ring has a one degree angle 
to it. So once the ring is in place, turn it a quarter turn 
counterclockwise and then begin to tighten by turning clockwise. 
The ring is seated when it is slightly recessed from the front plate. 
This ensures that the accessories will gently slide into place. Do 
not over tighten.  
 
 
 
Figure 23. Window replacement tool and replacing locking ring 
Replacing the Source 
The typical lifetime of the infrared source is three years. So it may 
need to be replaced on occasion. You should contact your local service 
representative if you are not sure whether or not the source should be 
replaced.  
NOTE 
To perform this procedure, you will need: 
•  3/32 ball driver 
•  Phillips screw driver 
 
Before starting this procedure, turn off your Cary 630 FTIR. Unplug 
the system from the main power source and from the USB connection 
to the computer. Also, remove any sampling accessory from the Cary 
630 FTIR so you are left with just the main unit.  
CAUTION 
Removing the sampling accessory will expose the ZnSe or KBr optical windows. 
Avoid touching or placing any object near the optical windows.